Polarity reversal procedure

from Wikipedia, the free encyclopedia

The polarity reversal method is a method from electrical measurement technology for interference suppression in complex impedance measurements with a fixed measurement frequency.

With this method, three measurements are made to determine the impedance of a test object at the measuring frequency:

  1. without source
  2. with source at measuring frequency
  3. with source at measuring frequency, 180 ° phase shifted
Figure 1: Vector diagram for the polarity reversal procedure

Figure 1 shows the complex vector diagram of a voltage measurement with a current fed in on a device under test using the polarity reversal method. Where:

  • a voltage caused by interferers measured with the source switched off,
  • the voltage measured at 0 ° phase shifted source,
  • the voltage measured with the source 180 ° out of phase.

The voltage sought for the impedance determination is calculated from the amounts of the measured voltages according to the following formula:

The impedance can then be calculated using the following formula:

When using the polarity reversal procedure, it is important to meet the following criteria:

  • The disturbance may only occur at one frequency.
  • The disturbance must be stable.

See also

credentials

  1. Power systems with nominal alternating voltages above 1 kV , from VDE standard 0101, VDE classification VDE 0101 , DIN number DIN VDE 0101 , edition 2000-01