Phase contrast microscopy

from Wikipedia, the free encyclopedia
The single cell Ammonia tepida in the phase contrast method

The phase contrast method (from Greek φασις, phasis "appearance" and Latin contra stare "to stand in opposition") is an imaging method in light microscopy . This makes use of the fact that, in addition to the amplitude , the phase of light waves also changes when they pass through a medium, depending on its refractive index . In this way, it is possible to directly image structures that have only a low intrinsic contrast and would only be visible with bright field microscopy with artificial coloring.

Phase contrast microscopy is widespread in the examination of biological samples in which the object parts differ only slightly in their light absorption but in their refractive index. The phase contrast method was developed in 1932 by the Dutch physicist Frits Zernike and introduced into microscopic practice in 1941 by the Jena Carl Zeiss works . In 1953, Zernike received the Nobel Prize in Physics for his discovery .

Contrast in light microscopy

Amplitude and phase object

The visibility of details in an image depends on the resolution and contrast of the image. In particular, the light microscopy of biological objects is impaired in many cases by the poor contrast of the images generated.

Amplitude and phase objects

Amplitude objects

When viewing in bright field, only areas in the image that absorb light to a certain extent in comparison to the surroundings appear clearly visible. Either through general light absorption (darkening) or through spectrally specific light absorption (natural colors). Such objects are also called amplitude objects.

Phase objects

More or less clear and colorless objects have little effect on the amplitude of the incident light and only appear visible if, as an exception, they have a very strong difference in the refractive index compared to the surroundings. As a rule, such objects have a very poor contrast, since the human eye cannot perceive phase differences as such. Since such objects, in contrast to amplitude objects, do not produce any significant difference in the amplitude of the incident light, but only a difference in phase, they are also called phase objects.

Methods of increasing the contrast

There are several methods of increasing the contrast. For example, individual objects or their components are specifically colored with dyes . Another method is that illumination to reduce by the iris of the illumination apparatus (the condenser narrowed). As a result, parts of the object with different refractive indices are displayed with greater contrast. However, this method has the serious disadvantage that the resolution of the image is greatly reduced, because the resolution of the microscope is dependent on the numerical aperture of the object illumination. In addition, it is difficult to visualize living biological material using artificial coloring, since many cells perish as a result of the coloring agent.

Principle of the phase contrast method

The phase contrast method from Frits Zernike uses differences in the refractive index and the thickness of the object to generate a light-dark contrast without significantly reducing the illumination aperture and thus the resolution of the microscope. Since light propagates at different speeds in media with different refractive indices, there is a phase difference when passing through an object that is optically denser than its surroundings compared to the light that does not pass through this object.

In order to be able to represent this phase shift in brightness differences, a phase ring in the objective and a ring diaphragm in the condenser lens are built into a phase contrast microscope . The ring diaphragm limits the incidence of light on the sample to a certain angle of incidence. If there is no specimen under the microscope, all of the background light falling through the ring diaphragm hits the phase ring. This consists of a material that attenuates the light and at the same time shifts its phase by 90 °.

However, if a preparation, for example a sample with transparent cells, is brought into the beam path, the light is partially deflected at the structures of the cells. In contrast to the undiffracted light, however, this diffracted object light predominantly does not run through the phase ring and is accordingly not influenced by it. However, the diffraction in the sample also causes a phase shift depending on the refractive index.

In the image plane, there is now interference between background and object light. In order to achieve the greatest possible contrast, the phase shift of the background light in the phase ring must be set so that the background light weakens the object light as much as possible when it interferes with the object light. The phase ring in the lens must therefore be dimensioned approximately according to the most common refractive indices and thicknesses of the objects being viewed. As a result, the object now appears mostly dark against the background. This is known as positive phase contrast . In the case of objects with a particularly high refractive index ( e.g. endospores of bacteria ), the contrast changes and they are displayed brighter than the background.

In a method that is seldom used, the phase ring is dimensioned in such a way that the contrast is reversed even with conventional objects, i.e. they appear light on a darker background; this is known as negative phase contrast .

By using the phase contrast method, the resolution of a microscope is not improved beyond the diffraction limit .

Applications

Trichoderma harzianum with conidiophores
phase contrast microscopy

The phase contrast method is most often used in light microscopy of biological objects. In particular when observing cells that are almost invisible in a normal light microscope, high-contrast images result without the need for staining.

When pollen analysis of the phase-contrast method can also fine structures of the surface with the help of pollen grains are made visible.

Furthermore, the image formation in high-resolution transmission electron microscopy and in high-resolution brightfield scanning transmission electron microscopy is based on the effect of phase contrast - but that of the wave functions of the beam electrons.

See also

literature

  • F. Zernike: The phase contrast method in microscopic observation. In: Z. techn. Physics. 16, 1935, pp. 454-457.
  • A. Köhler, W. Loos: The phase contrast method and its applications in microscopy. In: The natural sciences. No. 29, 1941, pp. 49-61.
  • K. Michel: The representation of chromosomes by means of the phase contrast method. In: The natural sciences. No. 29, 1941, pp. 61-62.
  • K. Michel: phase contrast. In: Zeiss communications. No. 1, 1959, pp. 243-267.
  • Heinz Gundlach: 50 years ago: Nobel Prize for Frits Zernike (1888–1966) in physics for the development of the phase contrast method. In: Current Cell Biology. Volume 29, No. 2, 2003, p. 15